3000RS PRECISION RESISTANCE STANDARD
OPERATION MANUAL
Transmille Ltd. Page 5
Design Notes
The 3000RS is built using oil filled hermetically sealed ultra high precision foil
resistors from 1 Ohm to 100kOhm, and oil filled hermetically sealed ultra high
precision wire wound for the 1 /10 MOhm – no adjustable resistors are used
within the design of this reference.
Features
Functions of the terminal post connections :
3000RS Terminal Post Function
CURRENT HI Input for the current source from an ohmmeter
CURRENT LO Input for the current source from an ohmmeter
VOLTAGE HI Measurement point for a four-wire ohmmeter
VOLTAGE LO Measurement point for a four-wire ohmmeter
EARTH Connected to the case